Kuderna-Danish evaporative concentrators are used to concentrate trace amounts of sample dissolved in organic solvents. The Snyder column is designed to speed evaporation with minimal hold-up.

The column should be pre-wetted with the solvent to be used in the concentration in order to minimize initial sample loss. The flask is then filled with solvent to between 40 and 60 percent of capacity. The charged apparatus should then be mounted over a boiling water bath so that the water level is maintained just below the lower ground joint, bathing the bottom of the flask with steam. The concentrator tube will collect the final sample for further analysis.

Proper safety precautions should always be followed. If solvent vapors are allowed to escape, the entire apparatus should be placed in a well-ventilated fume hood. 2 Keck Clamps are included. This version uses Keck Clamps instead of hooks and springs.

Unit of Measure

Specifications

Item No.

N/A 3121-06

Item Description

N/A Evaporative, Kuderna-Danish, with Graduated Concentrator Tube, 500mL Flask Capacity, 10mL Tube Capacity, 0-1 x 0.1mL Tube Graduations x 2-10 x 1mL Subdiv., 600mm Overall Height

Flask Capacity

N/A 500 mL

Tube Capacity

N/A 10 mL

Tube Graduations x Subdivision

N/A 0-1 x 0.1mL x 2-10 x 1mL

Approximate Overall Height

N/A 600 mm

Pkg. Qty.

N/A 1